CathLabHub complications toolkitcathlabhub.com/toolkit/wire-fracture-entrapmentIn review, October 24, 2026

Recognize

The signs on the monitor, the image and the patient.

  • ImageThe wire does not move with withdrawal, or its tip stays put while the shaft stretches 1.
  • ImageA wire tip that no longer responds, or a radiopaque fragment left in the vessel after withdrawal 1.
  • ImageAn unraveled coil, a bird's nest of metal, sometimes reaching the aorta 1.
  • PatientIschemia if the fragment causes thrombus or obstructs flow.

Act now

In the order the operator does them. The first three come first, whatever else is happening.

  1. Do not pull hard; pulling unravels or fractures the wire 1.
  2. Advance a microcatheter or a balloon over the wire as far as possible and pull the wire gently 1.
  3. If that fails, inflate the balloon advanced as far distally as possible and pull gently again 1.
  4. If that fails, pass a second wire beside the trapped one and inflate a balloon alongside to free it 1.
  5. Fractured wire: image with IVUS for unraveling 1.
  6. Fragments that can be trapped: stent over them against the wall 12.
  7. Long fragments in a large vessel: remove with wire twirling or a snare 1.
  8. Small distal fragments: leave in place, with dual antiplatelet therapy 12.
  9. Unraveled into the aorta or a critical location: surgery 1.

Equipment

From the complications cart. Linked items open in the equipment directory.

Drugs

Adult doses as the cited source gives them. Check your formulary and the label.

Who to call

The escalation ladder, first call first.

  1. Second operator: to manage the second wire and balloon.
  2. Cardiac surgery: for an unraveled wire in the left main or the aorta, or a wire that cannot be freed 1.

The bailout step by step

Freeing an entrapped wire 1

  1. Stop pulling; keep slight tension only.
  2. Advance a microcatheter or balloon over the trapped wire as far as it will go.
  3. Hold the microcatheter forward and pull the wire gently through it.
  4. Still stuck: inflate a balloon at the distal end of the reachable segment and pull gently again.
  5. Still stuck: pass a second wire beside it, inflate a balloon next to the trapped wire, and try again.
  6. Still stuck: surgery; a last resort is to pull hard with a balloon or microcatheter in place, accepting a fracture 1.

Prevent

  • Straighten a looped wire tip before pulling it back through a stent 1.
  • Do not over rotate wires; replace a kinked wire before atherectomy 1.
  • Do not jail the radiopaque segment of a side branch wire 4.

After care and documentation

  • Image the final result; document any retained fragment and its location.
  • Dual antiplatelet therapy if a fragment stays 2.
  • Keep the wire for analysis and report the event through your device reporting process.

Practice

No emergency simulator case for wire fracture or entrapment yet. Browse the case library.

References

  1. Brilakis ES. Manual of Percutaneous Coronary Interventions. Elsevier; 2021. Chapter 27, Equipment loss and entrapment.
  2. Berg R, Lim MJ. Complications of percutaneous coronary interventions. In: Kern MJ, ed. The Interventional Cardiac Catheterization Handbook. 3rd ed. Saunders, Elsevier; 2013. Chapter 4.
  3. Brilakis ES. Manual of Percutaneous Coronary Interventions. Elsevier; 2021. Chapter 25, Acute vessel closure.
  4. Brilakis ES. Manual of Percutaneous Coronary Interventions. Elsevier; 2021. Chapter 16, Bifurcations.

Educational reference for cath lab teams, not medical advice and not a standard of care. Your local policy, the device instructions for use and current guidelines govern. Verify every dose before use. Printed from cathlabhub.com/toolkit/wire-fracture-entrapment. First version, in clinical review: an interventional cardiologist is checking every card before it is final. Every number cites its source on the card.